Automated Pressure Calibrator
CPC 6000
82 Mensor/WIKA Operating Instructions - CPC 6000
9.2.2.1 Tuning a Solenoid Valve Regulator Module
Figure 9.2.2.1 - Tune Screen, Solenoid Valve Regulator
9.2.2.1.1 Modes
[Standby]: Seals all outside connections.
[Measure]: Isolates the active sensor and measure/control port.
[Control]: Controls to displayed setpoint in % F.S.
[Vent]: Opens manifold to atmosphere.
Setpoint = Desired pressure in % F.S.
Reading = Measured pressure in % F.S.
Supply = Supply fine bleed valve.
Balance = Balance between supply and exhaust fine valves.
Exhaust = Exhaust fine bleed valve.
[SAVE]: Saves the supply and exhaust fine bleed valves to the regulator.
Typical supply and exhaust values will be 3.00 to 6.00.