Wika 702.03.100 Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
GB
WIKA operating instructions dierential pressure gauge, model 702.03.10016
2367781.03 06/2013 GB/D/F/E
2123649.0X
7. Options and accessories / 8. Maintenance
7.2 Panel mounting ange
Panel
Cutout
8. Maintenance
The instruments are maintenance-free.
The indicator and switching function should be checked once or twice every year.
For this the instrument must be disconnected from the process to check with a
pressure testing device.
Repairs must only be carried out by the manufacturer.
Sequence of operations to dismount the measuring instrument with a
running process
1. Open the pressure compensating valve
2. Close the shut-o valve for the ⊕ and ⊖ media chamber
3. Open the vent valve
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 71 72

Comentarios a estos manuales

Sin comentarios