25WIKA operating instructions, models CTH6300 and CTH6500
2079988.02 04/2013 GB/D
GB
6. Commissioning, operation
Conrm selection by pressing ENTER/MENU.
Now use the arrow keys ▲▼ to select the required calibration function.
Standard linearisation according to DIN IEC 60751 [oFF
Use the arrow keys ▲▼ to select CoFF. Press ENTER/MENU to conrm the required
setting.
Press ESC to return to the measuring mode.
Calibration by code oP1
Use the arrow keys ▲▼ to select oP1. Press ENTER/MENU to conrm the required
setting. A small 1 appears in the lower part of the display. Four characters follow it (hex
code / 0 ... F). The 4 characters can be changed by means of the arrow key up ▲. To go
to the next number press the arrow key down ▼.
If all 4 characters are entered as required, conrm by pressing ENTER/MENU. A small 2
appears and the following 4 characters can be changed as well.
Press ESC to return to the measuring mode.
When calling the oP1 by means of ENTER/MENU the oP1 function
(calibration by code) is activated even if the menu has been closed by
pressing ESC.
MENU
Unit Prob Lin2 CAL Chnl ArEA Lo6
CAL
oFF oP1 oP2