Pressure Balance High-Pressure GB
CPB 5000 HP
WIKA Operating Instructions Pressure Balance High-Pressure Version 2.3
9
2.3.4 How the cross-sectional surface responds to pressure .
At higher pressures, the effective cross-sectional surface changes due to the pressure load.
The ratio of the cross-section and prevailing pressure is linear within an initial approximation. It is
represented by the coefficient of expansion caused by pressure distortion (λ).
pressure Nominal1
pressure Nominal
+
=
λ
pressureTrue
Example:
Measuring point: 1000 bar
System with distortion coefficient: 10
–7
1/bar:
barbarpressureTrue 90.999
10001011
1000
7
=
+
=
Without the correction, measurements would all be "off" by 0.01 %.