18 WIKA operating instructions, model CPH6400
11313439.04 12/2012 GB/D
GB
4. Design and function
When the battery is nearly empty, the message "low BAT" appears in the display. With
0 % battery level, the equipment automatically switches itself o and must then be
recharged using the battery charger.
4.5.1 During charging
CAUTION!
The temperature range over which the Lithium-Ion battery can be
charged is 10 °C ... 45 °C. Charging the Lithium-Ion battery at tempera-
tures outside of this range may lead to heating or damage. In addition,
the performance of the Lithium-Ion battery can be aected and the
service life reduced.
4.5.2 Charging the Lithium-Ion batteries
WARNING!
In order to charge the Lithium-Ion battery, never use any device other
than that specied by WIKA. When the Lithium-Ion battery is used in
devices other than the specied device, the performance and service
life of the Lithium-Ion battery may be reduced, and, should the device
cause an abnormal current to ow, it may cause the Lithium-Ion battery
to become hot, explode or ignite and result in serious injury.
CAUTION!
The temperature range over which the Lithium-Ion battery can be
discharged is -10 °C ... +60 °C. Use of the Lithium-Ion battery outside of
this temperature range may aect the performance of the battery or may
reduce its service life.