C
Universal Pressure Transmitter UniTrans
Operation of Transmitters with Display
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / BRD
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
64
6.2 Key Functions
6.3 The Programming Mode
The transmitter can be pro
g
rammed before or after installation.
The ke
y
pad is activated and the device can be pro
g
rammed b
y
simultaneousl
y
press-
in
g
the "esc" and "ok" ke
y
s (for 2 sec.). This method is used to access the main
menus. Each main menu has one or more sub-menus and each sub-menu, ma
y
have
its own sub-menus.
Button Functions
Main Menu Sub-menu Edit Functions
back to the previous
menu option
back to the previous
menu option
increase value
forward to next menu
option
forward to next menu
option
decrease value
back to value displa
y
without savin
g
back to main menu
without savin
g
back without savin
g
to the sub-menu to the edit functions save value
activate ke
y
pad (push simultaneousl
y
; 2 s)
The keypad becomes inactive after 10 min. of disuse. All settings will de-
fault to previously stored values. Only settings that have been confirmed
with the "OK" function are stored.
A change in the starting measurement (zero point) has no effect on the
measurement span. Likewise, a change in the span has no effect on the
starting measurement.
An error signal occurs when the zero point or span settings fall outside of
the sensors nominal pressure range during calibration with pressure.
Nothing is saved.