Pneumatic High-Speed Pressure Controller CPC3000
WIKA Operating Instruction Pneumatic High-Speed Pressure Controller ∙ Version 1.152
11498171.01 08/2009 GB
8.5.4 DPI 510 Commands Emulated
Command Data Response/Function
Q Return data in output format
A Make the A channel active
B Make the B channel active
M Go to local
R Select active range
S Select preset units
U Set units
D Select output data
F Select function
N Set output format
I Set interrupt conditions
C Select control mode
P Set the control setpoint
/ Setpoint ratio NOT EMULATED
* Setpoint preset NOT EMULATED
W Set the stable time
@ Error Report
V Set rate setpoint
J Rate setpoint preset NOT EMULATED
O Set zero offset
Q (baro) Get baro reading (secondary address 9)
A (baro) Select absolute
G (baro) Select gauge
DCL Device clear
GTL Go to local
IFC Interface clear
8.5.5 IEEE 488.2 Commands
Command Data Response/Function
*IDN? Return identification string
*RST Reset to known state (DEFAULT+PSI)
*TST? Returns OK
*OPC Operation completed
*WAI Returns operation completed state
*CLS Clear status and error queue
*ESE Enable status event