Hand Spindle Pump
CPP1000-M/-L
WIKA Operating Instructions Hand Spindle Pump18
If the blind plug is removed completely, a small
cylindrical body, made of red plastic, is visible.
This plastic body is to reduce the internal volume
of the pump if a shut-o valve or ne adjustment
valve is not used / mounted.
Remove the red cylindrical body using tweezers
or blow it out through the open test connection
using compressed air.
Now screw in the shut-o valve or the ne adjust-
ment valve.
Pull tight the valve using a at wrench 27 mm.
This parts are left over and may be depolluted:
blind plug
red cylindrical body