10. Troubleshooting
Defect Possible reason Remedy
No signal output Failure of power supply or
wiring interrupted
Check power supply and
wiring; replace defective
components
Transmitter miswired Check wiring; if necessary
rectify it
No pressure (only 3-wire)
Check tailpipes
Electronic defect through incor-
rect supply voltage or external
voltage
Return instrument to manufac-
turer for repair
Steady signal, despite
pressure changes
Pressure entry blocked Check tailpipes and pressure
entry bore, if necessary clean it
carefully
Supply voltage connected the
wrong way round
(I = approx. 4.5 mA)
Check the clamping and, if
necessary, change terminals 1
and 2 around
Transmitter failure Return instrument to manufac-
turer for repair
Steady and exces-
sively high signal,
despite pressure
changes
Bridge between terminals 5
and 6 (I = approx. 9.5 mA) was
not removed
Open the cable hood and
remove the bridge between the
terminals 5 and 6
Electronic defect through incor-
rect supply voltage or external
voltage
Return instrument to manufac-
turer for repair
Full span reading too
low
Supply voltage too low Adjust supply voltage
Load impedance too high Consider permissible max. load
Zero point shifted Readjust the zero point (see
chapter 9)
Zero signal too low/
too high
Zero point shifted Readjust the zero point (see
chapter 9)
11297264 05/2009 GB/D/F/E
WIKA Operating Instructions intelliGAUGE family 19
10. Troubleshooting