WIKA Operating instructions / Betriebsanleitung, WUC-1x
Failure Possible cause Procedure
No/False output signal Incorrectly wired (e.g. Connected as 2-wire
instead of 3-wire system)
Follow pin assignment (see Instrument Label /
Operating Instructions)
Abnormal output signal or
Abnormal zero point signal
Zero point set wrongly Adjust zero point correctly (see chapter 8); a
suciently accurate current/volt meter should
be used
Abnormal zero point signal Medium or ambient temperature too high/
too low
Control the internal temperature of the instru-
ment within the permissible range; observe the
allowable temperature error (see Operating
Instructions)
Diaphragm is damaged, e.g. through impact,
abrasive/agressive media; corrosion of
diaphragm/pressure connector.
Replace instrument
Signal span dropping o/too small Diaphragm is damaged, e.g. through impact,
abrasive/agressive media; corrosion of
diaphragm/pressure connector
Contact the manufacturer and replace the
instrument
Signal span too small Power supply too high/too low Correct the power supply in line with the
Operating Instructions
Mechanical overload through overpressure Re-calibrate the instrument *)
In case of unjustied reclamation we charge the reclamation handling expenses.
*) Make sure that after the setting the unit is working properly. In case the error continues to exist send in the instrument for reparation (or replace the unit).
10. Trouble shooting GB
If the problem persists, contact our sales department.
USA, Canada: If the problem continues, contact WIKA or an authorized agent for assistance. If the pressure transmit-
ter must be returned obtain an RMA (return material authorization) number and shipping instructions from the place of
purchase. Be sure to include detailed information about the problem. Pressure transmitters received by WIKA without a valid
RMA number will not be accepted.
Process material certicate (Contamination declaration for returned goods)
Purge / clean dismounted instruments before returning them in order to protect our employees and the environment from any
hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been submitted and fully lled-in. This
Return Form contains information on all materials with which the instrument has come into contact, either through installa-
tion, test purposes, or cleaning. You can nd the Product Return Form on our internet site (www.wika.com).