Wika 70-8xx Manual de usuario Pagina 12

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 11
GB
12
11472988.02 04/2013 GB/D/F/E
WIKA operating instructions expansion thermometer, model 70
7. Indicator check
Indicator checks should only be carried out in comparison with a more accurate
instrument or, if possible, with a calibrated instrument. The temperature during the check
must remain constant.
Fluctuating temperatures can lead to reading errors caused by the dierent response
times of the sensors.
With thermometers without thermowells, with a full and correct insertion length, a
minimum waiting time of 5 minutes must be maintained for temperature equalisation.
Temperature checks using thermometers with sensors assembled with thermowells can,
in many cases, only result in a reference temperature at the instrument under test, due to
permanent heat dissipation caused by the thermowell. In temperature applications tted
with static checking thermometers (e.g. pipeline systems), permanent osets in reference
temperatures can also be caused by the measuring path. Ambient temperatures around
the indicator case which dier substantially from room temperature can lead to steady
indication errors with constant ambient temperatures and to varying indication errors with
uctuating ambient temperatures.
Permanent indication errors caused by the ageing of the measuring system can be
ignored, as they only account for a fraction of the indication accuracy.
8. Zero adjustment
WARNING!
Any interference with or modication to the instrument will invalidate the
warranty!
A zero adjustment may only be carried out by the manufacturer or in adequately equipped
workshops by qualied persons.
The electrical connection must only be made by qualied skilled personnel.
Connection of the switch via terminal box
Conductor cross-section max. 1.5 mm
2
The terminal assignment is stated on the connection label of the thermometer
7. Indicator check / 8. Zero adjustment
Vista de pagina 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 59 60

Comentarios a estos manuales

Sin comentarios