Specications A-10 IS-20
Data sheet PE 81.60 PE 81.50
Design standard intrinsically safe
Pressure ranges bar 0 ... 6 to 0 ... 60 0 ... 6 to 0 ... 60
Outputs mA 4 ... 20 4 ... 20 (line transformer)
Medium temperature °C -30 ... +100 -20 ... +80
Ambient temperature °C -30 ... +100 -20 ... +80
Wetted parts stainless steel stainless steel
Power supply U
B DC 8 V < UB ≤ 30 V 10 V < UB ≤ 30 V
Maximum load R
A Ohm RA ≤ (UB - 8 V) / 0.02 A RA ≤ (UB - 10 V) / 0.02 A
Accuracy
BFSL % of span ≤ 0.5 ≤ 0.25
Compensated temperature
range
°C 0 ... +80 °C 0 ... +80 °C
Wiring details,
2-wire
11592011.04 06/2014 GB/D/F/I
GB
WIKA operating instructions dierential pressure gauge models 712.15.100, 732.15.100
13
9. Transmitter for working pressure indication
9. Transmitter for working pressure indication (optional)
Standard version model A-10
or Ex version model IS-20
Transmitter for working
pressure indication
The corresponding operating instructions are included in the delivery
of each dierential pressure gauge with integrated transmitter for
working pressure indication.
Non hazardous
area
Hazardous
(classied) area
The transmitters for the working pressure
are screwed in sideways, on the left side of
the minus measurement chamber and can,
if necessary, be retrotted on-site.
Pressure connection for Transmitter:
G 1/4 (male)