Specications A-10 IS-20
Data sheet PE 81.60 PE 81.50
Design standard intrinsically safe
Pressure ranges bar 0 ... 6 to 0 ... 60 0 ... 6 to 0 ... 60
Outputs mA 4 ... 20 4 ... 20 (line transformer)
Medium temperature °C -30 ... +100 -20 ... +80
Ambient temperature °C -30 ... +100 -20 ... +80
Wetted parts stainless steel stainless steel
Power supply U
B DC 8 V < UB ≤ 30 V 10 V < UB ≤ 30 V
Maximum load R
A Ohm RA ≤ (UB - 8 V) / 0.02 A RA ≤ (UB - 10 V) / 0.02 A
Accuracy
BFSL % of span ≤ 0.5 ≤ 0.25
Compensated temperature
range
°C 0 ... +80 °C 0 ... +80 °C
Wiring details,
2-wire
11592011.04 06/2014 GB/D/F/I
GB
WIKA operating instructions dierential pressure gauge models 712.15.100, 732.15.100
13
9. Transmitter for working pressure indication
9. Transmitter for working pressure indication (optional)
Standard version model A-10
or Ex version model IS-20
Transmitter for working
pressure indication
The corresponding operating instructions are included in the delivery
of each dierential pressure gauge with integrated transmitter for
working pressure indication.
Non hazardous
area
Hazardous
(classied) area
The transmitters for the working pressure
are screwed in sideways, on the left side of
the minus measurement chamber and can,
if necessary, be retrotted on-site.
Pressure connection for Transmitter:
G 1/4 (male)
Vista de pagina 12