Wika CPH65I0 Manual de usuario Pagina 38

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 100
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 37
38 WIKA operating instructions, model CPH65I0
14061681.02 03/2013 GB/D
GB
6. Commissioning, operation
3. Open the vent key on the pump and zero the calibrator. Then close the vent after
resetting the calibrator.
4. The top of the display will indicate "CLOSE".
5. Apply pressure to the pump slowly until the pressure switch opens.
In the switch test mode the display update rate is increased to help
capture changing pressure inputs. Even with this enhanced sample rate
the device under test should be pressurized slowly in order to ensure
accurate readings.
6. Once the switch is open, "OPEN" will be displayed; bleed the pump slowly until the
pressure switch closes.
7. In the top display it will now read "SW OPENED AT" and give you the pressure at
which the switch opened is displayed.
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 ... 99 100

Comentarios a estos manuales

Sin comentarios