15WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
GB
6. Commissioning, operation
Sealing of the process connection
For further information on sealings see WIKA data sheet AC 09.08 or under www.wika.com.
Rotatability of the case
The case of the model DG-10-E can be rotated through 300°, thus the
display can be read from almost any position.
Do not turn the case forcibly past the stop point.
Parallel threads Tapered threads
Correct sealing of the process connections with
parallel threads at the sealing face must be
made using suitable at gaskets, sealing rings or
WIKA prole sealings.
For sealing process connections with tapered
threads, the sealing must be made in the threads
using additional sealing material, e.g. PTFE tape
(EN 837-2).
300°
per EN 837 per DIN 3852-E NPT, R and PT