6. Commissioning, operation
6.4.8 Pump limit
To prevent overpressuring for sensitive test items, a maximum pressure (pump limit) can
be set. In this mode, use the arrow keys to set the maximum pressure.
6.4.9 HART™ Resistor
An internal 250 Ω HART™ resistor can be enabled when the calibrator is operated
in the "mA measure-24V" mode. This allows a HART Communicator to be connected
across the mA terminals and eliminates the need for adding an external resistor.
When the HART™ resistor is enabled, the maximum load driving
capability is 750 Ω.