3.9 Safety-related maximum values
(only valid for devices to ATEX category 3G)
Model GCS-1-A Ex nL IIC T4, T5 X Ex nA IIC T4, T5 X
Supply voltage: DC 14 ... 24 V DC 14 ... 24 V
Current limitation I
i
: 30 mA -
Power limitation P
i:
1 W 1 W
Internal capacitance C
i:
0,1 µF
1)
-
Internal inductance L
i:
1,2 mH
2)
-
Model GCS-1-G Ex nL IIC T4, T5 X Ex nA IIC T4, T5 X
Supply voltage: DC 14 ... 24 V DC 14 ... 24 V
Current limitation I
i
: 30 mA -
Power limitation P
i:
1 W 1 W
Internal capacitance C
i:
0,1 µF
1)
-
Internal inductance L
i:
1,2 mH
2)
-
Max. output voltage U
o:
DC 12 V -
Max. output current I
o:
35 mA -
External capacitance C
o:
4,122 µF -
External inductance L
o:
23,4 mH -
Model GCS-1-F Ex nL IIC T4, T5 X Ex nA IIC T4, T5 X
Supply voltage: DC 14 ... 24 V DC 14 ... 24 V
Current limitation I
i
: 30 mA -
Power limitation P
i:
1 W 1 W
Internal capacitance C
i:
0,1 µF
1)
-
Internal inductance L
i:
1,2 mH
2)
-
Max. output voltage U
o:
DC 12 V -
Max. output current I
o:
35 mA -
External capacitance C
o:
4,122 µF -
External inductance L
o:
23,4 mH -
1) 56 pF / cable length in m
2) 0.1 µH / cable length in m
3. Specications