Wika MHC-1 Manual de usuario Pagina 15

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 84
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 14
15WIKA operating instructions pressure transmitter, model MHC-1
14035467.01 08/2012 GB/D/F/E
GB
6. Commissioning, operation
6. Commissioning, operation
WARNING!
Prior to commissioning, the pressure transmitter must be subjected to a visual inspection.
Only use the pressure transmitter if it is in perfect condition with respect to safety.
6.1 Making the mechanical connection
The sealing faces at the instrument have to be undamaged and
clean.
When screwing the instrument in, the force required to do this must
not be applied through the casing, but only through the spanner ats
provided for this purpose and using a suitable tool.
The correct torque depends on the dimensions of the process
connection and the gasket used (form/material).
Maximum tightening torque
Process connection Tightening torque
G ½ A DIN 3852-E 60 Nm
M14 x 1.5 60 Nm
7/16-10 UNF-2A 20 Nm
¼ NPT 34 Nm
When screwing in, do not cross the threads.
For information on tapped holes and welding sockets, see Technical information IN 00.14 at
www.wika.com.
Spanner ats