WIKA operating instructions model TR30 per ATEX 17
GB
minimum maximum
-50 ... -20 -50 ... +250
-40 ... -20 -40 ... +250
0 ... +20 0 ... +250
+10 ... +30 +10 ... +250
+20 ... +50 +20 ... +250
+50 ... +230 +50 ... +250
Measuring range in °C
-50 °C
-40 °C
0 °C
+10 °C
+20 °C
+50 °C
Range of possible upper limits to measuring range in °C
-50 0 +100 +250
Start of measuring
range
Diagram for measuring range
With conguration to customer specications, the measuring range will be shown clearly
on the instrument label. The conguration and the choice of parameters is described in the
menu-driven conguration software. Changes to the conguration must be noted on the
label using a water-resistant felt-tip pen.
Possible combinations of the lower and upper limits of the measuring range
The upper limit of the measuring range is dependent on the respective lower limit of the
measuring range.
For clarity, this dependency has been shown in 50 °C steps in the examples in these
diagrams. The conguration software checks the required measuring range and will only
accept permissible values. Intermediate values are congurable; the smallest increment is
0.1 °C.
7. Conguration of model TR30-W
Please note:
The measuring range of the thermometer is limited by the application range of the
measuring element, not by the adjustable range of the transmitter.
Without neck tube -50 … +150 °C, with neck tube -50 … +250 °C.