14068347.01 06/2014 GB/D
GB
6 Commissioning, operation
4. Attach the push-on cap or display and operating unit and screw the case head cover
tight down to the stop.
- Ensure that the case head is tightly closed.
5. Perform a mounting correction.
- Without display, see chapter 8.1 “Performing a mounting correction (oset)”
- Via HART
®
, see chapter 8.2 “Conguring the HART
interface”
- With display, see chapter 9.5 “Mounting correction (oset)”
6.2.3 Pin assignments
Angular connector DIN 175301-803 A
1
2
3
+ 1
2
Shield GND
Circular connector M12 x 1 (4-pin)
4
3
1
2
+ 1
3
Shield 4
With shielded connection cable, connect the cable shield to the internal
grounding screw.
Pin assignmentOutlet for connection cable
Cable gland
Negative power supply terminal
Process connection
Ground screw, inside (GND)
Positive power supply terminal +
Spring-loaded terminal