Wika UPT-21 Manual de usuario Pagina 28

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 120
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 27
28
WIKA operating instructions process transmitter, model UPT-2x
14068347.01 06/2014 GB/D
GB
7. Display and operating unit, model DI-PT-U
7.5 Settingtheadditionaldisplay
The additional display can indicate the following values:
Measuredvalues
Pressure
Level
Current
PV percent
Sensor temperature
Dragpointervalues
P
min
/ P
max
PV
min
/ PV
max
T
min
/ T
max
Furtherdata
TAG short (max. 8 capital letters and gures)
TAG long (max. 32 alphanumeric characters)
Blank (disables the additional display)
1. Open the operating menu with [↵].
Select “Display” and conrm with [↵].
2. Select Add. display” and conrm with [↵].
3. Select value and conrm with [↵].
Additional display indicates the selected value.
Vista de pagina 27