3. Installation / 4. Dierential pressure gauge
11248832.02
6
Measuring cell Adjustable measuring ranges
from to
140 mbar 0 ... 40 mbar - 0 ... 140 mbar
280 mbar 0 ... 80 mbar - 0 ... 280 mbar
560 mbar 0 ... 160 mbar - 0 ... 560 mbar
1130 mbar 0 ... 320 mbar - 0 ... 1130 mbar
2300 mbar 0 ... 650 mbar - 0 ... 2300 mbar
4000 mbar 0 ... 1150 mbar - 0 ... 4000 mbar
11265191.07 12/2012 GB/D/F/I
WIKA operating instructions dierential pressure gauge models 712.15.160, 732.15.160
GB
4. Dierential pressure gauge
The measuring range of the dierential pressure gauge can, depending on the particular
measurement system used, be adjusted within the limits given in the previous specications
table). Ideally, this adjustment should be made on a test bench, though it can also be carried
out at the measuring point using a hand test pump.
Measuring range limits
Turn clockwise: reduce measuring range
Turn anticlockwise: expand measuring range
Panel cut-out
Panel
Adjustable measuring span
Adjustable measuring span
Allen key
(included in delivery)
Option
Panel mounting
Cover cap for adjustable
measuring span