Wika CPC2000 Manual de usuario Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 132
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
23WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
6.5 SETUP settings (SETUP)
The SETUP menu items are made up of six submenus.
RANGE: Setting of the pressure range (max. 100 % of the measuring span)
UNIT: Selection of the pressure unit (Pa, kPa, hPa, bar, mbar, psi, inH
2
O, inHg, mmHg,
Torr)
STEPS: Setting of the step size in x %. Settable in the range 1 ... 50 %
MODE: Selection of the operating mode
Control (CTRL): Regulates to the given set point
Measure (MEAS): Measurement of dierential and relative pressure
Automatic (AUTO): Carries out a stored test routine
LANGUAGE: Selection of the displayed language
SETTINGS: Here additional instrument settings can be made
Navigation within the menu:
SELECT: Acts to conrm the relevant setting
Left/Right arrow: Setting of the active decimal point
Up/Down arrow: Change between the dierent submenu points.
6.5.1 The menu points
With the MENU button and with the
/ button, press the up and down until one is
at the required menu point. With the SELECT button, one reaches the desired submenu.
The following items are available in the submenu:
6.5.1.1 Range
Under the RANGE menu point the controller‘s nal control value can be changed for the
control.
Vista de pagina 22