30 WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
9. Faults / 10. Dismounting, return and disposal
9. Faults
Faults Causes Measures
Instrument is not working,
the display is dark.
No voltage supply
■
Check whether the plug is
correctly in the socket
■
Battery is empty and must
rst be recharged
Set pressure has not
been reached, the pump
is running continuously
System is not fully sealed;
to large hose diameter or
hose length
■
Slide hose on correctly;
eliminate any leaks
■
Hose diameter max. 6 mm
(see chapter 3 "Specica-
tions“)
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the
low-pressure controller must be shut down immediately, and it must be
ensured that pressure and/or signal are no longer present, and it must be
prevented from being inadvertently put back into service.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
10.2 “Return”.
10. Dismounting, return and disposal
WARNING!
Residual media on the dismounted low-pressure controller can result in a
risk to persons, the environment and equipment. Take sucient precau-
tionary measures.
10.1 Dismounting
Only disconnect test and calibration installations once the system has been depressu-
rised!
When disassembling the instrument the following must be followed:
1. Ensure that there is no positive or negative gauge pressure in the instrument and that
all parts of the instrument are at room temperature.
2. Switch the instrument o using the ON/OFF switch