Wika LH-20 Manual de usuario Pagina 25

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 24
25
8 Supplement
WIKA Operating instructions - Submersible pressure transmitter model LH-20
Voltage supply
Operating voltage 4 … 20 mA
Ʋ Deviation < 0.2 % 8 … 30 V DC
Ʋ Deviation < 0.1 % 9.6 … 30 V DC
Operating voltage 4 … 20 mA/HART 9.6 … 30 V DC
Permissible residual ripple
Ʋ < 100 Hz U
ss
< 1 V
Ʋ 100 Hz … 10 kHz U
ss
< 10 mV
Reverse voltage protection Integrated
load max. ≤ (U
B
-10 V)/0.022 A
Integrated overvoltage protection
Nominal leakage current (8/20 µs) 5 kA
Min. response time < 25 ns
Electrical protective measures
Protection transmitter IP 68 (30 bar)
Overvoltage category III
Protection class III
CE conformity
ATEX 94/9/EG (optional)
4 … 20 mA:
EMC directive 2004/108/EG, EN 61326 Emission (group 1, class B) and interference resistance
(industrial area)
4 … 20 mA/HART:
EMC directive 2004/108/EG, EN 61326 Emission (group 1, class A) and interference resistance
(industrial area)
Warning:
according to EN 55011 or CISPR 11
This is an equipment of class A for emitted interference and is designed for use in
industrial environment. In other environments such as e.g. in living or business areas, this
equipment can interfere other instruments.
Vista de pagina 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 59 60

Comentarios a estos manuales

Sin comentarios