Inuence of the product or ambient temperature
Applies to instruments with digital signal output (HART, Probus PA, Foundation Fieldbus) as well
as to instruments with analogue current output 4 20 mA. Specications refer to the set span.
Turn down (TD) = nominal measuring range/set span.
Thermal change zero zignal, reference temperature 20 °C (68 °F):
- In the compensated temperature range
0 +100 °C (+32 +212 °F)
< 0.05 %/10 K x TD
- Outside the compensated temperature
range
typ. < 0.05 %/10 K x TD
Applies also to instruments with analogue 4 20 mA current output and refers to the set span.
Thermal change, current output < 0.15 % at -40 +80 °C (-40 +176 °F)
Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)
Applies to digital interfaces (HART, Probus PA, Foundation Fieldbus) as well as to analogue
current output 4 20 mA. Specications refer to the set span. Turn down (TD) = nominal
measuring range/set span.
Long-term drift of the zero signal < (0.1 % x TD)/year
Ambient conditions
Ambient, storage and transport temperature
- Standard version -40 +80 °C (-40 +176 °F)
- Standard version with indicating and
adjustment module
-15 +70 °C (+5 +158 °F)
- Connection G1 A front-ush according to
EHEDG
-10 +80 °C (+14 +176 °F)
- Fitting G1 A front-ush according to
EHEDG with indicating and adjustment
module
-10 +70 °C (+14 +158 °F)
- Version for oxygen applications
13)
-40 +60 °C (-40 +140 °F)
- Version for oxygen applications with
indicating and adjustment module
14)
-15 +60 °C (+5 +140 °F)
Process conditions
The specications are used as an overview. For pressure stage and product temperature the
specications on the type plate are applicable.
Product temperature standard, depending on the seal
15)
- FKM (VP2/A) -20 +105 °C (-4 +221 °F)
- EPDM (A+P 75.5/KW75F) -40 +105 °C (-40 +221 °F)
- NBR (COG) -20 +105 °C (-4 +221 °F)
13)
Up to 60 °C (140 °F).
14)
Up to 60 °C (140 °F).
15)
Version for oxygen applications up to 60 °C (140 °F).
46 Process pressure transmitter IPT-1* Vers. 3.0 - metalic sensor 4 20 mA/HART