Wika R-1 Manual de usuario Pagina 11

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 92
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 10
11WIKA operating instructions pressure transmitter, model R-1
14026284.01 11/2012 GB/D/F/E
GB
3.2 Output signals
Signal type Signal
Current (2-wire) 4 ... 20 mA
Voltage (3-wire) DC 0 ... 10 V
DC 1 ... 5 V
Ratiometric (3-wire) DC 0.5 ... 4.5 V
Other output signals available on request
Load in Ω
Current output (2-wire): ≤ (power supply - 7 V) / 0.02 A
Voltage output (3-wire): > maximum output signal / 1 mA
Ratiometric output (3-wire): > maximum output signal / 1 mA
3.3 Voltage supply
Power supply
The power supply depends on the selected output signal
4 ... 20 mA: DC 7 ...30 V
DC 1 ... 5 V: DC 8 ...30 V
DC 0 ... 10 V: DC 14 ... 30 V
DC 0.5 ... 4.5 V: DC 4.5 ... 5.5 V
3. Specications