WIKA Additional Instructions Pressure Transmitter with CANopen Interface D-20-9 / D-21-9 26
2330h Auto-Operational
Object description
Through this object the user can set the device to enter the OPERATIONAL-Mode automatically after boot up.
Entry description
Access rw
PDO Mapping No
Data Type UNSIGNED 8
Default Value 00h
Data Range 00h ... 01h
Changed parameters are not automatically stored, but they have to be stored via object 1010h Store
Parameters.
Furthermore the device can be reset to the factory settings via object 1011h Restore Default Parameters.