49WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
ENTER
ENTER
7.3.5 CALIBRATION mode (calibration of a pressure transmitter)
7. Commissioning, operation
Deviation
Deviation
Deviation
No. of test item
and No. of test
point
No. of test item
and No. of test
point
No. of test item
and No. of test
point
Generate the specied
setpoint in acc. with the
reference display (establish
a pressure-free condition/
atmosphere) and with
<--
record the measured values
of this test point
Generate the specied
setpoint in acc. with the
reference display using the
pressure generator
Record the measured values
of this test point
(with
back to previous
test point)
2. Test point
x. Test point
1. Test point
(e.g. 0 bar)