
40 Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
A? <sp>n.nnnnne+nn<cr><lf> Returns the A channel pressure reading.
AR? <sp>n.nnnnne+nn<cr><lf> Returns the A channel rate.
BARO? <sp>n.nnnnne+nn<cr><lf> Returns reading from barometric sensor
B? <sp>n.nnnnne+nn<cr><lf> Returns the B channel pressure reading.
BR? <sp>n.nnnnne+nn<cr><lf> Returns the B channel rate.
CALDISABLE yes,no Sets whether or not calibration of the ac-
tive channel is disabled.
CALDISABLE? <sp>(yes or no)<cr><lf> Returns whether or not calibration of the
active channel is disabled.
CERR none Clears the error queue.
CHAN A or B Sets the active channel on the instrument.
CHAN? <sp>A or B or D or
Baro<cr><lf>
Returns which channel is active.
CHAN D none Sets the active channel to the Delta chan-
nel.
CHANFUNC Press, peak, rate, baro Sets the alternate function mode of the
active channel.
CHANFUNC? <sp>CCCCC…<cr><lf> Returns the alternate function mode of the
channel.
DECPT? <sp>n<c><lf> Returns the number of decimal points for
the active channel (see RESOLUTION).
DEFAULT none Sets the default values.
DELTAFUNC A-B none Sets the delta to be the result of the chan-
nel A reading - channel B reading.
(Chan D must be in the active channel).
DELTAFUNC B-A none Sets the delta to be the result of the chan-
nel B reading - channel A reading.
(Chan D must be in the active channel).
DELTAFUNC A+B none Sets the delta to be the result of the chan-
nel A reading + channel B reading.
(Chan D must be in the active channel).
DELTAFUNC? <sp>{A-B or B-A}<cr><lf> Returns delta function as one of the above
options. (Chan D must be in the active
channel).
Comentarios a estos manuales