58 Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
7.1.3.8.4 General SCPI WIKA Commands
STATus
:OPERation
:CONDition? Returns an integer value representing instru-
ment status that can be decoded.
Bit 0: Zeroing active.
Bit 1: Control Setpoint has not been reached.
Bit 2: Reserved 0.
Bit 3: Reserved 0.
Bit 4: Measuring. The instrument is actively
measuring.
MEASure
[:PRESsure] [z]? Returns the pressure in the current units from
the specified sensor, or active sensor if [z] is
omitted.
:TEMPerature[z]? Returns the temperature in deg. C. from the
specified sensor, or active sensor if [z] is omit-
ted.
:RATE[z]? Returns the pressure rate in the current units
from the specified sensor, or active sensor if [z]
is omitted.
:BAROmetric? Returns the barometric pressure in the current
unit, if a barometer sensor is installed.
CALibration
:MODE? Returns 1 if the sensor has been calibrated or 0
if not.
:DATE? Returns the stored calibration date.
:DATE <i,i,i> Sets the calibration date (YYYY,MM,DD).
:ZERO? Returns current zero point correction in bar.
:ZERO <n> Sets the zero so the currently measured
pressure matches the commanded value in the
current unit. Limited to +/-5% FS.
ZERO:RUN Vents the instrument and sets zero so the
currently measured pressure = 0.