52 Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
UNITS_TABLE not emulated
UNITS_TABLE? UNITS_TABLE? lf not emulated
WINDOW window sp
<value>lf
Sets the filter window for pressure readings where
<value> is a value in the current measurement
units within the range of the instrument.
WINDOW? window?lf Returns the filter window setting for the pressure
reading.
The instrument returns <value><cr><lf> where
<value> is a value in the current units within the
range of the instrument.
ZERO zero sp <value>lf This command requires that the PWZ or the PW
password must be issued before the first occur-
rence of the zero command. The zero command
sets the DPG zero offset to <value> in the current
units where <value> is a value between +/- 17 psi.
ZERO? ZERO?lf Returns the current zero calibration offset where
<value> is current zero value in current measure-
ment units.
7.1.3.8.2 UNITS Command Syntax for DPG 2100 Measurement Units
Table 7.1.3.8.2 - UNITS Command Syntax for DPG 2100 Measurement Units
N Description Output Format Type
1 pounds per square inch PSI English
2 inches of mercury @ 0°C INHG English
3 inches of mercury @ 60°F INHG English
4 inches of water @ 4°C INH2O English
5 inches of water @ 20°C INH2O English
6 inches of water @ 60°F INH2O English
7 feet of water @ 4°C FTH2O English
8 feet of water @ 20°C FTH2O English