Wika T12 Manual de usuario Pagina 20

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 19
GB
3211517.0707/2010GB/D
20 WIKAoperatinginstructionstemperaturetransmitterT12
6.2.2 Sensors
Schematic representation / conguration
6. Commissioning, operation
Resistance thermometer (RTD) and resistance sensor
Itispossibletoconnectanresistancethermometer(e.g.to
DINEN60751)ina2-,3-or4-wireconnectionmethod.Congurethe
inputofthetransmittertomatchwiththeactualmethodofconnection
used.Otherwiseyouwillnotfullyexploitthepossibilitiesofconnec-
tionleadcompensationand,asaresult,possiblycauseadditional
measuringerrors(seechapter„6.2Conguration“).
Thermocouples (TC)
Makesurethatthethermocoupleisconnectedwiththecorrect
polarity.Iftheleadbetweenthethermocoupleandthetransmitter
needstobelengthened,onlyusethermalorcompensationcablein
accordancewiththetypeofthermocoupleconnected.
Conguretheinputofthetransmitterappropriatelyforthethermo-
coupletypeandthecoldjunctionactuallyused,otherwisemeasure-
menterrorsmaybecaused(seechapter"6.2Conguration").
Shouldthecoldjunctioncompensationbeoperatedwith
anexternalresistancethermometer(2-wireconnection),
connectthistoterminalsand.
Thermocouple
Resistancethermometer/
linearresistance
in
4-wire3-wire2-wire
CJCwith
external
Pt100/Ni100
1)
Input
mVsensor
4 ... 20 mA loop
CJC
internal
1)Connectsensor(Pt100/Ni100)forexternalcoldjunctioncompensationbetweenterminal1and4.
Vista de pagina 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 71 72

Comentarios a estos manuales

Sin comentarios