14 WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
3. Specications
20
3
42
17
126
Ø 6
Ø 17,5
G ½ B
Ø 40
M24 x 1,5
29,5
Specications CPT6000 reference pressure sensor
Pressure connection
≤ 1000 bar: G ½ B; {various connection adapters on request}
> 1000 bar: M16 x 1.5 female, with sealing cone
Material Stainless steel for wetted parts (for measuring ranges
> 25 bar ... ≤ 1000 bar additional Elgiloy
®
)
Internal transmission uid
Synthetic oil (only for measuring ranges up to 25 bar)
{Halocarbon oil for oxygen variants}
4)
Permissible temperature ranges
Medium
°C -20 ... +80
Storage
°C -40 ... +85
Case Stainless steel
Ingress protection IP 65 (with cable connected)
Weight g approx. 230
CE conformity
Pressure equipment directive 97/23/EC; module A
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and
interference immunity (industrial application)
4) For oxygen versions, the medium temperature must not exceed 60 °C.
{} Items in curved brackets are optional extras for an additional price.
For further specications see WIKA data sheet CT 15.01 and the order documentation.