Wika CPH6000 Manual de usuario Pagina 52

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 132
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 51
52 WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
Customer-specic unit; with respect to bar
(input via numeric keypad)
Display resolution in operating mode
(back with )
Select and conrm (standard units) via
Short info:
XXX
Current cursor position; alter via
◁▷ ◀▶
Parameter selection from list or menu via
0.00 Parameter input via numeric keypad
ENTER
Input conrmation
CLEAR
Clear entry
In order to switch the instrument into "SWITCH-TEST" mode, follow the instructions on
the previous page.
The switch-test function is not suitable for electronic switches (e.g.
PNP- or NPN switches), but is only for mechanical, potential-free
switches.
The following is a more detailed explanation of point "2. Preparation for SWITCH-
TEST"
a) Voltage supply for test item (on/o) [If no external supply is required for the
test item, "OFF" should be selected to conserve energy]
b) Unit and resolution (submenu)
7. Commissioning, operation
The calculated measured values of the two switching points and the
hysteresis can be reset by pressing the "0" key.
Vista de pagina 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 131 132

Comentarios a estos manuales

Sin comentarios